Product Description:

Stryker iBur 2.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-012-020D

Product Classification:

Class II

Date Initiated: July 24, 2024
Date Posted: September 4, 2024
Recall Number: Z-2724-2024
Event ID: 95040
Reason for Recall:

Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

Status: Completed
Product Quantity: 1226 units
Code Information:

GTIN: 07613327501261 Lot Numbers: 22237017 22241017 22334017 23085017 23179017 23213017 24003017 24015017

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated