Stryker Corporation: Medical Device Recall in 2024 - (Recall #: Z-2731-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Stryker iBur 3.0mm Precision Match Head, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-107-530
Class II
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
GTIN: 07613327501193 Lot Numbers: 22228017 22257017 22269017 22276017 23085017 23095017 23157017 23157027 23179017 23179027 23198017 23213017 23221017 23299017 23345017 23345027
Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.
Voluntary: Firm initiated