Stryker Corporation: Medical Device Recall in 2025 - (Recall #: Z-1469-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Otopore Cylinder Standard outer ear dressings, Catalog Numbers 5400-010-000 and 5400-010-000ITL

Product Classification:

Class II

Date Initiated: February 25, 2025
Date Posted: April 9, 2025
Recall Number: Z-1469-2025
Event ID: 96393
Reason for Recall:

There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.

Status: Ongoing
Product Quantity: 182,344 total units
Code Information:

5400-010-000 UDI-DI: 07613327301663, Lot Numbers:2023011112, 2023020334, 2023072413, 2024032615; 5400-010-000ITL UDI-DI: 07613327359510, Lot Numbers: 2023022132 2023112213 2024020515 2024032013

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated