Stryker Endoscopy: Medical Device Recall in 2014 - (Recall #: Z-0638-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

PKG, BABCOCK FORCEPS, PADDLE, P/N 0250080250. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Product Classification:

Class II

Date Initiated: November 17, 2014
Date Posted: December 24, 2014
Recall Number: Z-0638-2015
Event ID: 69838
Reason for Recall:

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Status: Terminated
Product Quantity: 180,573 units total
Code Information:

847693 940794 941851 942139 945976 1040104 1042237 1042684 1042713 1142907 1144319 1145981 1244080 1246130 1340545 1341999 1343014 1442503

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated