Stryker Endoscopy: Medical Device Recall in 2014 - (Recall #: Z-0722-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

PKG, 3.5MM CANNULA AUTO VALVE, STOPCOCK, SHORT, P/N 0250080634 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Product Classification:

Class II

Date Initiated: November 17, 2014
Date Posted: December 24, 2014
Recall Number: Z-0722-2015
Event ID: 69838
Reason for Recall:

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods.

Status: Terminated
Product Quantity: 180,573 units total
Code Information:

38042 41043 41849 41875 42240 42502 44647 46566 47092 47503 47788 48285 49179 49691 52200 52203 53377 54246 56194 60879 63395 73215 74024 74683 75672 76832 77049 83778 84741 86162

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated