Stryker Endoscopy: Medical Device Recall in 2014 - (Recall #: Z-0730-2015)
See the recall detail below. You can also see other recalls from the same firm in 2014.
PKG, ASPIRATION NEEDLE, 33CM, P/N 0250080702 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Class II
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
34052 35336 36936 37411 38777 42671 43115 43965 46127 46128 47098 47968 48904 51714 55844 58062 60075 62764 64137 64566 66123 67425 69336 69922 71368 72053 72916 74022 75105 76697 78532 79394 81390 81784 82272 82724 85034 86632 87673 88122 88852
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Voluntary: Firm initiated