Stryker Endoscopy: Medical Device Recall in 2014 - (Recall #: Z-0758-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 20CM, P/N 0250282000 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Product Classification:

Class II

Date Initiated: November 17, 2014
Date Posted: December 24, 2014
Recall Number: Z-0758-2015
Event ID: 69838
Reason for Recall:

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Status: Terminated
Product Quantity: 180,573 units total
Code Information:

342970 645484 743722 747490 747891 840091 945208 945712 946001 1040121 1041472 1042482 1042952 1046585 1140928 1143074 1143076 1143121 1145963 1146536 1241511 1244708 1246167 1341512 1343689 1344271 1344682 1440414

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated