Stryker Endoscopy: Medical Device Recall in 2014 - (Recall #: Z-1872-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.

Product Classification:

Class II

Date Initiated: May 19, 2014
Date Posted: July 2, 2014
Recall Number: Z-1872-2014
Event ID: 68405
Reason for Recall:

STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.

Status: Terminated
Product Quantity: 1992 trays
Code Information:

Part Number 272-700-000: Lot number: 20476.1 23541 23837 26403 26404 26405 26676 27009 27518 28341 30248 34355 35932 36756 39705 05J034834 06con05774 06CON06360 06CON06846 06con06869 06con06960 06con08587 25067-1-12 26882-1-1 26882-1-11 26882-1-12 26882-1-13 26882-1-14 35200-1-1 36610-1-1 38820-1-1 39078-1-1 40504-2-1 40675-1-1 41231-1-1 41827-1-1 42085-1-1 42375-1-1 42527-1-1 42789-1-1 43103-1-1 43589-1-1 43716-1-1 43716-1-3 43716-1-4 44568-1-10 44568-1-11 44568-1-2 44568-1-3 44568-1-4 44568-1-6 44568-1-7 44568-1-8 44568-1-9 46227-1-1 47731-1-1 48595-1-1 48596-1-1 48597-1-1 48598-1-1 48599-1-1 48600-1-1 50101-1-1 50628-1-1 50828-1-1 51237-1-1 52439-1-1 55035-1-1 55422-1-1 55462-1-1 56456-1-1 R471611 SS15329.

Distribution Pattern:

Worldwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated