Stryker Endoscopy: Medical Device Recall in 2014 - (Recall #: Z-2080-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Disposable StrykeFlow Tip, Stryker Disposable Suction/I irrigator Tip; Sterile, Rx only, Latex Free. Stryker Suction Irrigation system, used to provide a clean and clear operative site through the evacuation and aspiration of irrigant solution. for endoscopic use.

Product Classification:

Class II

Date Initiated: June 20, 2014
Date Posted: July 30, 2014
Recall Number: Z-2080-2014
Event ID: 68624
Reason for Recall:

Internal review found that six lots of Disposable StrykeFlow Tips were shipped with a potentially deformed blister pouch which could impact sterility of the product.

Status: Terminated
Product Quantity: 496 boxes (6 tips per box)
Code Information:

P/N 0250070505; Lot numbers: 14021FG2, 13337FG2, 13343FG2, 13347FG2, 14007FG2, 14016FG2.

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Internationally tAustralia, Canada, France, Greece, Netherlands, United kingdom and South Korea.

Voluntary or Mandated:

Voluntary: Firm initiated