Stryker Endoscopy: Medical Device Recall in 2016 - (Recall #: Z-1566-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator; High pressure units (HPU) (40L and 45L). Model Numbers: 620-040-610, 620-040-611, 620-040-613, 620-040-614, 620-040-600, 620-040-601, 620-040-602, 620-040-604, 105-210-684 The Pneumo Sure XL High Flow Insufflator is a device that serves to create a cavity by insufflating CO2 during diagnostics and/or therapeutic laparoscopy. The insufflator consists of a console, that is intended to be connected to a CO2 gas source, and has a tubeset inserted to transport CO2 gas from the insufflator to the connecting trocar.

Product Classification:

Class II

Date Initiated: March 28, 2016
Date Posted: May 4, 2016
Recall Number: Z-1566-2016
Event ID: 73674
Reason for Recall:

When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display, immediately suspending insufflation.

Status: Terminated
Product Quantity: 489
Code Information:

HPUs affected lot numbers 0005051337-000505052795; PneumoSure insufflators affected serial numbers 1510CE0342-1601CE0581

Distribution Pattern:

Worldwide distribution. US nationwide including Puerto Rico, China, Hong Kong, India, Japan, Korea, Romania, Vietnam, Singapore, Taiwan, Netherlands, Peru, Australia, Canada, Mexico, Argentina, Chile, Poland, South Africa, and Guatemala.

Voluntary or Mandated:

Voluntary: Firm initiated