Stryker Instruments Div. of Stryker Corporation: Medical Device Recall in 2013 - (Recall #: Z-0280-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Acetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0206-522-000 is an accessory to the Advanced Cement Mixer (ACM). The Acetabular Pressurizer Tiplet P/N 0206-522-012 is sealed in a two pouches, sterilized, and labeled as part number 0206-522-000. This Acetabular Pressurizer is an accessory used to pressurize the cement applied to the prosthesis to be implanted during the surgery. The Acetabular Pressurizer is an injection molded component made with resin material DOW HDPD 10462N which is a rigid plastic once it is molded.

Product Classification:

Class II

Date Initiated: November 1, 2013
Date Posted: November 20, 2013
Recall Number: Z-0280-2014
Event ID: 66444
Reason for Recall:

On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be placed in the pouch to be sealed, the manufacturing operator noticed that some assemblies were cracked. As a result of the injection pressure of the cement gun, a cracked tiplet may release from the Hub during cement applica

Status: Terminated
Product Quantity: 87 boxes (522 each)
Code Information:

Part number: 0206-522-000 and lots: 08183012, 08310012, 09237012, 10243012, 11109012, 11353012, and 12216012

Distribution Pattern:

Worldwide distribution: US (nationwide) including states of: TX, FL, IL, IA, ND, WI, VA, and NJ.; and countries of: CANADA, POLAND and SWITZERLAND.

Voluntary or Mandated:

Voluntary: Firm initiated