Stryker Instruments Div. of Stryker Corporation: Medical Device Recall in 2013 - (Recall #: Z-1497-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Stryker Instruments 0400-820-000S3 T4 Toga w/Inside Shoulder Ties, Sterile, O; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Product Classification:

Class II

Date Initiated: May 23, 2013
Date Posted: June 12, 2013
Recall Number: Z-1497-2013
Event ID: 65321
Reason for Recall:

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Status: Terminated
Product Quantity: 3000 units (300 boxes)
Code Information:

1007145, 1101004, 1107487

Distribution Pattern:

USA Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated