Stryker Instruments Div. of Stryker Corporation: Medical Device Recall in 2016 - (Recall #: Z-1378-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone cutting tips which are a part of Strykers Neuro Spine Ear, Nose and Throat (ENT) Sonopet Portfolio. They are removable attachments intended to be used with the Sonopet Ultrasonic Handpiece.
Class II
The Spetzler Claw Tip was packaged with the incorrect Tip Sleeve Assembly. The Tip Sleeve Assembly included in the packaging is the incorrect size. Risk to Health: If the Tip Sleeve Assembly does not fit the Spetzler Claw" Tip, a surgical delay may occur while an alternate Tip Sleeve Assembly is retrieved.
5450-800-311 Lot Number 0545410 5450-800-313 Lot Number 0555170
Distributed in the states of AZ, CA, CO, CT, IL, IN, MD, MI, MN, NY, PA, SC, TN, WA, WI, and VA, and in the NETHERLANDS.
Voluntary: Firm initiated