Stryker Medical Division of Stryker Corporation: Medical Device Recall in 2013 - (Recall #: Z-0629-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls.

Product Classification:

Class II

Date Initiated: December 18, 2012
Date Posted: January 9, 2013
Recall Number: Z-0629-2013
Event ID: 63602
Reason for Recall:

Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.

Status: Terminated
Product Quantity: 1004
Code Information:

Model 3005S3EX, Serial Number 120415087 - 120515468,

Distribution Pattern:

Worldwide Distribution - Nationwide and Canada

Voluntary or Mandated:

Voluntary: Firm initiated