Stryker Neurovascular: Medical Device Recall in 2020 - (Recall #: Z-1621-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Stryker Neurovascular, AXS Vecta 74 Aspiration Catheter, Ref. Catalog No. INC-11597-115, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459241(17)211129(10)00000-00 Stryker Neurovascular, AXS Vecta 74 Aspiration Catheter, Ref. Catalog No. INC-11597-125, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459258(17)211129(10)00000-00 Stryker Neurovascular, AXS Vecta 74 Aspiration Catheter, Ref. Catalog No. INC-11597-132, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459265(17)211129(10)00000-00 Product Usage: The AXS Vecta Aspiration System, including the AXS Vecta Aspiration Catheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral-M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.

Product Classification:

Class II

Date Initiated: April 10, 2019
Date Posted: April 8, 2020
Recall Number: Z-1621-2020
Event ID: 84780
Reason for Recall:

The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.

Status: Terminated
Product Quantity: 1405 total devices
Code Information:

Catalog # INC-11597-115: (Lot # 13728-01) Catalog # INC-11597-125: (Lot # 13730-01) Catalog # INC-11597-132: (Lot # 13732-01)

Distribution Pattern:

US Nationwide distribution in the states of PA, FL, CA, IL, WI, AZ, AR, FL, TN, MI, MA, CA, NC, WV, DE, OH, NJ, MO, NY, SC, TX, AL, VA, UT, MD, GA, LA, CO, WA, KY, OK, RI, SD, MS, NE,IA, KS, DC. No US Govt. No Foreign countries.

Voluntary or Mandated:

Voluntary: Firm initiated