Superstat Corp: Medical Device Recall in 2012 - (Recall #: Z-0336-2013)
See the recall detail below. You can also see other recalls from the same firm in 2012.
Superstat Modified Collagen Hemostatic Sponge, 2x2 Pledget, 1.5%, Part# 9105-00. For use when hemostasis is desired along suture lines, in diffusely bleeding sites, in incisions, dissections, and around arterial or venous anastomoses. Indicated in oozing bleeding, either capillary or venous, in almost all types of incisions or wounds, and as an adjunct to conventional closure techniques
Class II
The recall was initiated because Superstat Corporation has confirmed that no documentation are available to verify that quarterly dose sterility audits were performed for the Superstat Modified Collagen Hemostatic Sponge.
Lot# SSC02, SSF02 .
Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, Thailand, Mexico, Canada, Pakistan, Israel, Greece, Serbia, Canda, Isle of Man, Nova Scotia, and Singapore.
Voluntary: Firm initiated