Surgical Theater Inc: Medical Device Recall in 2025 - (Recall #: Z-0322-2026)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
SyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SNP.1.12.0, imaging processing radiological system
Product Classification:
Class II
Date Initiated: September 16, 2025
Date Posted: November 5, 2025
Recall Number: Z-0322-2026
Event ID: 97715
Reason for Recall:
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
Status: Ongoing
Product Quantity: 14 units
Code Information:
UDI/DI 00860003884102
Distribution Pattern:
Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Voluntary or Mandated:
Voluntary: Firm initiated