Symbios Medical Products, LLC: Medical Device Recall in 2013 - (Recall #: Z-1457-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Part 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits GoBlock Kit 510472, Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.
Class I
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
Pump Part 510449, Lot 11-100271 - Lot 130100137 found in: GoBlock Kit 510472, Lot 11-100385 to 13-100354
US Nationwide distribution.
Voluntary: Firm initiated