Symbios Medical Products, LLC: Medical Device Recall in 2013 - (Recall #: Z-1459-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Part 510490 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml per side)( 6mL/hr ),... found in the following kits GoBlock Kit 510491, Disposable Single Infusion Pump for drug delivery
Class I
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
Pump Part 510490, Lot 11-100379- Lot 12-101132 found in: GoBlock Kit 510491, Lot 11-100601 to 13-100705
US Nationwide distribution.
Voluntary: Firm initiated