Symbios Medical Products, LLC: Medical Device Recall in 2013 - (Recall #: Z-1461-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Part 510613 Disposable Infusion Pump, ...Elastomeric Pump 300mL (150 mL side)( 6mL/hr ),... found in the following kits GoBlock Kit 510612, Disposable Infusion Pump for drug delivery. Symbios GOBlock System is a disposable, self contained infusion system.
Class I
Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the
Pump Part 510613 , Lot 11-100815 - Lot 12-10101060 found in: GoBlock Kit 510612, Lot 11-100817 to 13-101475
US Nationwide distribution.
Voluntary: Firm initiated