Synthes, Inc.: Medical Device Recall in 2014 - (Recall #: Z-2674-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
DePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4 Level / 64 mm part 04.613.364 lot 8143905
Class II
DePuy Synthes is initiating a voluntary medical device recall of certain lots of the TI Vectra Plates, which is a part of the DePuy Synthes Vectra Anterior Cervical Plate Family of Systems. The Vectra Systems are intended for anterior screw fixation to the cervical spine (C2-C7). It was discovered in certain lots, that the clips are potentially missing from the TI Vectra Plates resulting in the i
TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4 Level / 64 mm part 04.613.364 lot 8143905
US Distribution including the states of IL, SC, VA, TX, MA, NC, LA, CA, AL, CO, NE, OR and KS., and Internationally to Canada.
Voluntary: Firm initiated