Product Description:
LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments
Product Classification:
Class II
Date Initiated: August 7, 2017
Date Posted: February 28, 2018
Recall Number: Z-0669-2018
Event ID: 79021
Reason for Recall:
May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.
Status: Terminated
Product Quantity: 567 units
Code Information:
Lot numbers FT00227 9972971 H028400 H161606 H288454
Distribution Pattern:
Distribution to US, Canada, and Germany.
Voluntary or Mandated:
Voluntary: Firm initiated