Synthes USA HQ, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0393-2014)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Synthes Material Mix (Adjustable Cervical Distractor-Right, Matrix Distractor Rack, T-PAL Spacer Remover, Implant Inserter) Synthes Material Mix: Adjustable Cervical Distractor-Right is used to distract spine segment. Matrix Distractor Rack is used to distract the interbody disc space. T-PAL Spacer Remover is used to remove T-PAL spacer. Implant Inserter is used to insert the ProDisc implant.
Class II
One material batch used for production of the affected parts of the Synthes Material Mix were mixed in with an incorrect material with a lower than specified hardness and corrosion resistance.
part no. 396.396 (lot nos. T989765 and T989780), part no. 03.632.087 (lotT987270), part no. 03.812.005 (lot nos. T987734, T988931, and T987623), and part no. 03.820.129 (lot no. T988206).
Worldwided Distribution: USA (nationwide) including IL, MO, TX, WA and Internationally to Canada and Switzerland.
Voluntary: Firm initiated