Synthes USA HQ, Inc.: Medical Device Recall in 2013 - (Recall #: Z-1238-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Synthes Radial Head Prosthesis: Trial Radial Head Elbow joint prosthesis: Synthes Radial Head Prosthesis System is a two-piece modular system comprised of titanium alloy stem and cobalt chrome components with an integral screw and side-loading application to allow in-situ assembly.

Product Classification:

Class II

Date Initiated: February 22, 2013
Date Posted: May 15, 2013
Recall Number: Z-1238-2013
Event ID: 64771
Reason for Recall:

The Trial Head may come loose from the implant stem during manipulation of the arm during surgery. To date, no adverse events have been reported related to this issue and this recall is not being initiated as a result of adverse events.

Status: Terminated
Product Quantity: 417
Code Information:

All Lots of Part Numbers: 03.402.018, 03.402.020, 03.402.022, 03.402.024, 03.402.026, 03.402.028, 03.402.218, 03.402.220, 03.402.222, 03.402.224, 03.402.226, 03.402.228, 03.402.418, 03.402.420, 03.402.422, 03.402.424, 03.402.426, 03.402.428, 03.402.618, 03.402.620, 03.402.622, 03.402.624, 03.402.626, and 03.402.628.

Distribution Pattern:

Nationwide Distribution: USA including states of: IL, IN, MA, MI, MO, NJ, NY, OH, OR, PA, SC, TX, VA, WA and WI.

Voluntary or Mandated:

Voluntary: Firm initiated