Synthes (USA) Products LLC: Medical Device Recall in 2016 - (Recall #: Z-0867-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Synthes VA Implant Rack for the Compact Distal Radius System Tray/Surgical/Instrument
Product Classification:
Class II
Date Initiated: November 29, 2016
Date Posted: December 28, 2016
Recall Number: Z-0867-2017
Event ID: 75828
Reason for Recall:
The screw rack plate was assembled incorrectly which could cause the screws in the screw rack to sit low and touch the surface beneath the screw rack.
Status: Terminated
Product Quantity: 31 units
Code Information:
Catalog ID (part # 60.119.012), Lot Number FL00528
Distribution Pattern:
Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated