Synthes (USA) Products LLC: Medical Device Recall in 2018 - (Recall #: Z-2245-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Part no. 690.375, Graphic Case F/6.5mm & 7.3mm Combined Cannulated Screw Set

Product Classification:

Class II

Date Initiated: March 5, 2018
Date Posted: June 27, 2018
Recall Number: Z-2245-2018
Event ID: 79754
Reason for Recall:

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

Status: Terminated
Product Quantity: 7
Code Information:

FL00860

Distribution Pattern:

Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.

Voluntary or Mandated:

Voluntary: Firm initiated