Product Description:

Part no. 690.379, Graphic Case F/6.5mm Cannulated Screw Set

Product Classification:

Class II

Date Initiated: March 5, 2018
Date Posted: June 27, 2018
Recall Number: Z-2246-2018
Event ID: 79754
Reason for Recall:

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

Status: Terminated
Product Quantity: 4
Code Information:

FL00860

Distribution Pattern:

Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.

Voluntary or Mandated:

Voluntary: Firm initiated