Sysmex America, Inc.: Medical Device Recall in 2012 - (Recall #: Z-0254-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Sysmex XE-2100 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #983-1341-1 The Sysmex XE-2100 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-2100 performs hematology analyses according to the RF/DC detection method, HydroDynamic Focusing (DC Detection), flow cytometry method (using a semiconductor laser), and SLS-hemoglobin method.

Product Classification:

Class II

Date Initiated: June 1, 2012
Date Posted: November 14, 2012
Recall Number: Z-0254-2013
Event ID: 62580
Reason for Recall:

There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

Status: Terminated
Product Quantity: 1,099 units
Code Information:

catalog #983-1341-1, all serial numbers

Distribution Pattern:

Nationwide distribution: USA including Puerto Rico.

Voluntary or Mandated:

Voluntary: Firm initiated