Sysmex America, Inc.: Medical Device Recall in 2012 - (Recall #: Z-0256-2013)
See the recall detail below. You can also see other recalls from the same firm in 2012.
Sysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #053-2311-4 The Sysmex XE-2100C is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
Class II
There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.
catalog ##053-2311-4, all serial numbers
Nationwide distribution: USA including Puerto Rico.
Voluntary: Firm initiated