Sysmex America Inc: Medical Device Recall in 2013 - (Recall #: Z-0100-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

pocH-100i; sysmex; Automated Hematology Analyzer; 7.3" x 13.8" x 18.1" / 30.8 lbs.; Sysmex Corporation; 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073, Japan; Sysmex America, Inc; One Nelson C. White Pkwy, Mundeline, IL 60060, U.S.A. The Sysmex pocH-100i Automated Hematology Analyzer is an automated cell counter intended for in vitro diagnostic use in CLIA non-waived clinical laboratories (not for Point of Care use in a CLIA waived laboratory). The Complete Blood Cel (CBC) count test parameters include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, and NEUT%/#, LYMPH %/#, MIXED %/# (EO, BSO, MONO).

Product Classification:

Class II

Date Initiated: May 15, 2013
Date Posted: November 6, 2013
Recall Number: Z-0100-2014
Event ID: 65722
Reason for Recall:

The data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number.

Status: Terminated
Product Quantity: 193 units
Code Information:

Product Code: 023-1861-8; Serial Numbers: B1677 - B1826, F1827 - F2161, F2178 - F2198, F2216 - F2396, F2411 - F2452, F2464 - F2472, F2495 - F2523, F2533, F2550 - F2560, F2592 - F2601, F2614 - F2635, F2640 - F2648, F2662 - F2679, F2710 - F2853, F2855 - F2871, F2878 - F2922, F2945, F2946, F2962 - F2970, F2998 - F3008, F3072 - F3080, F3083 - F3098

Distribution Pattern:

Worldwide distribution: US (Nationwide) including Puerto Rico and states of: AK, AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, SD, TN, TX, UT, VA, VT, WA, WI, and WY; and country of Canada.

Voluntary or Mandated:

Voluntary: Firm initiated