Tandem Diabetes Care, Inc.: Medical Device Recall in 2024 - (Recall #: Z-1609-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology
Class I
During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.
version 2.7 or later /UDI: 00850006613274
US nationwide distribution including Puerto Rico.
Voluntary: Firm initiated