Tandem Diabetes Care, Inc.: Medical Device Recall in 2025 - (Recall #: Z-2621-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

Product Classification:

Class II

Date Initiated: July 3, 2025
Date Posted: October 1, 2025
Recall Number: Z-2621-2025
Event ID: 97264
Reason for Recall:

an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.

Status: Ongoing
Product Quantity: 19
Code Information:

Tandem t:slim Mobile App, Android Software Version 2.9.1 UDI-DI code: 00389152470641 Serial Numbers / Cell Phone OS: 1163923 Google Pixel 8 Pro husky [15 35] 1164430 samsung SM-S901E r0qxxx [15 35] 1167131 samsung SM-A346E a34xdxx [14 34] 1299238 samsung SM-G780G r8qxx [13 33] 1300180 samsung SM-S928B e3qxxx [15 35] 1300180 samsung SM-S928B e3qxxx [14 34] 1329242 Google Pixel 6 Pro raven [15 35] 1381313 samsung SM-G780G r8qxx [13 33] 1381315 Google Pixel 9 tokay [16 36] 1381537 samsung SM-A546E a54xnsxx [14 34] 1381570 Xiaomi 2407FPN8EG rothko_global [15 35] 1391808 samsung SM-S711B r11sxxx [14 34] 1391892 samsung SM-N980F c1sxx [13 33] 1392863 samsung SM-S926B e2sxxx [15 35] 1392953 samsung SM-S916B dm2qxxx [15 35] 1393040 samsung SM-S901E r0qxxx [15 35] 90615975 realme RMX3241 RMX3241 [13 33] 90754587 samsung SM-S916B dm2qxxx [15 35] 901144584 samsung SM-S901E r0qxxx [14 34]

Distribution Pattern:

International distribution to the country of Israel.

Voluntary or Mandated:

Voluntary: Firm initiated