Tango3 LLC: Medical Device Recall in 2012 - (Recall #: Z-2138-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC, 132 Citizens Blvd., Simpsonville, KY 40067. Models are WOZ-250-PE, WOZ-240-SS, WOZ-180-SS. The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility.

Product Classification:

Class II

Date Initiated: March 26, 2012
Date Posted: August 15, 2012
Recall Number: Z-2138-2012
Event ID: 62514
Reason for Recall:

During an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either not reviewed or received proper clearance from FDA. There was no assurance that all systems in the field were working properly and operating at a minimum of 0.2ppm of ozone.

Status: Terminated
Product Quantity: 44 units
Code Information:

Model WOZ -250-PE, Serial #s: 1031, 1032, 1033, 1035, 1036, 1038, 1039, 1042, 1043, 1044, 1045, 1046, 1047, 1048, 1049, 1050, 1051, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1063, 1065, 1066, 1067, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076 & 1077. Model WOZ-180-SS, Serial #s: 1034, 1037, 1040 & 1064. Model WOZ-240-SS, Serial #s: 1041 & 1052.

Distribution Pattern:

Product was shipped to the following states: CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, ME, NH, NJ, NM, OH, PA, RI, TX and VA.

Voluntary or Mandated:

Voluntary: Firm initiated