Tecan US, Inc.: Medical Device Recall in 2015 - (Recall #: Z-0991-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Infinite 200 The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The infinite 200 has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384-well microplates conforming to the ANSI/SBS standards.

Product Classification:

Class III

Date Initiated: November 10, 2014
Date Posted: January 21, 2015
Recall Number: Z-0991-2015
Event ID: 69739
Reason for Recall:

Incorrect lumi firmware version installed (E.027 instead of V2.00)

Status: Terminated
Product Quantity: 8 units
Code Information:

Material number: 30016056, Lot numbers: 708004532, 810007441, 906000758.

Distribution Pattern:

Worldwide Distribution - USA including MD and Internationally to Hong Kong, South Korea and the UK.

Voluntary or Mandated:

Voluntary: Firm initiated