Tecan US, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2923-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.
Class II
Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.
In USA RE62119 Lot # LCL129 exp. date: 2017-11-30, RE62119 Lot # LCL133 exp. date: 2019-03-31, and RE62111 Lot # LCL130 exp. date: 2018-03-31. OUS Lot 131, Lot 132
Worldwide distribution; US distribution to states of: GA, IA, LA, MA, MN, OR, PA, SC, and UT; and countries of: Germany, Finland, Switzerland, France, Austria, Denmark, Poland, Norway, Belgium, Netherlands, Italy, Canada, South Africa, Russia, and Japan.
Voluntary: Firm initiated