Technidata S.A.: Medical Device Recall in 2013 - (Recall #: Z-1743-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: Laboratory Information System (L.I.S)

Product Classification:

Class III

Date Initiated: June 20, 2013
Date Posted: July 24, 2013
Recall Number: Z-1743-2013
Event ID: 65652
Reason for Recall:

A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and

Status: Terminated
Product Quantity: 1 site
Code Information:

All TDHistology/Cytology product versions.

Distribution Pattern:

US Distribution CA only.

Voluntary or Mandated:

Voluntary: Firm initiated