TECHNO-PATH MANUFACTURING LTD.: Medical Device Recall in 2025 - (Recall #: Z-0502-2026)
See the recall detail below. You can also see other recalls from the same firm in 2025.
Multichem ID-B. Model Number: SR103B. Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory testing procedures for the determination of anti-HIV-1 IgG, anti-HBc IgG, anti-HTLV-I IgG, anti-HCV IgG, HBs antigen and anti-Treponema pallidum IgG.
Class II
Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform.
Model Number: SR103B. UDI: (01)05391523442312(240)SR103B(17)261031(10) SB161024P. Lot Number: SB161024P. Expiration Date: 31-10-2026.
US distribution to CA, CO, CT, FL, HI, IL, LA, MI, MN, NV, NY, OR, SC, TN, TX, WI, WY.
Voluntary: Firm initiated