Teleflex Medical: Medical Device Recall in 2016 - (Recall #: Z-0886-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Auto Endo5 Hem-o-lok Automatic Clip Applier, Medium Large Polymer Clips, Rx Only, Sterile, REF 543965 Product Usage: The Weck Auto Endo5 hem-o-lok ML automatic endoscopic ligating clip appliers are indicated for use as delivery devices for Hem-o-lok ML non-absorbable polymer ligating clips. These appliers are designed for use with 5/5.5mm cannulas. Hem-o-lok Ligating Clips are intended for use in procedures involving ligation of vessels or tissue structures.

Product Classification:

Class II

Date Initiated: February 2, 2016
Date Posted: March 2, 2016
Recall Number: Z-0886-2016
Event ID: 73181
Reason for Recall:

The clip in the applier may be missing a boss. A boss is the part of the clip that is used to hold the clip within the applier jaws prior to application.

Status: Terminated
Product Quantity: 543 ea.
Code Information:

73J1500065, 73J1500066, 73J1500280, 73J1500440, 73J1500565, 73J1500566, 73K1500193

Distribution Pattern:

Worldwide Distribution in the state of CA, MI, NC, PA, TX, VA, and the countries of Canada, Korea & Belgium

Voluntary or Mandated:

Voluntary: Firm initiated