Teleflex Medical: Medical Device Recall in 2017 - (Recall #: Z-0059-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Teleflex(R) Percuvance(TM) Percutaneous Surgical System, Components: (a) 5 mm Atraumatic Johans Grasper, Catalog Number PCVJG5 (b) 5 mm Maryland Dissector, Catalog Number PCVMD5 (c) 5 mm Hem-o-lok Clip Applier, Catalog Number PCVHCA5 (d) 5 mm Traumatic Gripper Grasper, Catalog Number PCVGG5 (e) 5 mm Scissors, Catalog Number PCVSC5 The Percuvance Percutaneous Surgical System is a family of instruments used to perform laparoscopic procedures. Without the need for additional insertion conduits (trocars), the 2.9mm insulated Shafts are inserted percutaneously with the assistance of an Introducer needle tool tip. The Shaft is then advanced retrograde through a 5mm, or larger, port to facilitate the extracorporeal attachment of interchangeable Tool Tips. These interchangeable tools can be used to grasp, retract, manipulate, cut, coagulate and deliver Hem-o-lok¿ ligating clips to soft tissue. Each Percuvance System Handle includes an integrated male cautery adapter that can be connected to most monopolar electrosurgical units (ESU) via a standard 4mm monopolar cable.

Product Classification:

Class II

Date Initiated: October 18, 2017
Date Posted: November 8, 2017
Recall Number: Z-0059-2018
Event ID: 78340
Reason for Recall:

Teleflex is recalling these products because the jaws of the tool tip may break, causing a fragment of the tool tip to fall off during use.

Status: Terminated
Product Quantity: 3904 units
Code Information:

(a) Catalog Number PCVJG5, Lot Numbers: 73E1700794, 73G1700271, 73G1700515, 73H1700081, 73H1700367 (b) Catalog Number PCVMD5, Lot Numbers: 73E1700796, 73G1700273, 73H1700083 (c) Catalog Number PCVHCA5, Lot Numbers: 73E1700798, 73H1700084 (d) Catalog Number PCVGG5, Lot Numbers: 73E1700793, 73G1700272, 73H1700082, 73H1700247 (e) Catalog Number PCVSC5, Lot Number: 73H1700569

Distribution Pattern:

Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia, Belgium, Hong Kong, Japan, S. Korea and Singapore.

Voluntary or Mandated:

Voluntary: Firm initiated