Terumo Cardiovascular Systems Corporation: Medical Device Recall in 2017 - (Recall #: Z-1201-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Hercules 360 Universal Stabilizing Arm, top mount, 360-degree rotation, reusable stainless steel
Class II
Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.
Serial Number Range 70194 through 70504
Terumo Cardiovascular has identified 16 domestic consignees and 30 international consignees .
Voluntary: Firm initiated