Terumo Cardiovascular Systems Corporation: Medical Device Recall in 2017 - (Recall #: Z-1232-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment
Class II
Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.
Catalog Number: 401-231, Product Description: Titan" Stabilizer Attachment , Lot Number: 052R Catalog Number 401-231u , Product Description TitanTM 360 Stabilizer Attachment , Lot Number: 020R, 024R, 026R, 029R, 030R,034R,035R,037R, 040 , 048R
The product was distributed domestically and also to foreign consignees. But not to government agency's or Canada.
Voluntary: Firm initiated