Thermedx LLC: Medical Device Recall in 2014 - (Recall #: Z-2555-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of used irrigant.
Class II
A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.
Model P4000 Serial numbers affected: 20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005
Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.
Voluntary: Firm initiated