Thermo Fisher Scientific (Asheville) LLC: Medical Device Recall in 2019 - (Recall #: Z-1865-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05
Product Classification:
Class II
Date Initiated: April 2, 2019
Date Posted: July 3, 2019
Recall Number: Z-1865-2019
Event ID: 82903
Reason for Recall:
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Status: Terminated
Product Quantity: 10 units
Code Information:
Serial Numbers: 30024679S 300246796 300246797 300246798 300246799 300247393 300247394 30024739S 300247396 3002477S8
Distribution Pattern:
CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Voluntary or Mandated:
Voluntary: Firm initiated