Thermo Fisher Scientific (Asheville) LLC: Medical Device Recall in 2019 - (Recall #: Z-1869-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Thermo Scientific General Purpose Dynamic Bath Precision 20L GP Bath Model Number: TSGP20
Product Classification:
Class II
Date Initiated: April 2, 2019
Date Posted: July 3, 2019
Recall Number: Z-1869-2019
Event ID: 82903
Reason for Recall:
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Status: Terminated
Product Quantity: 4 units
Code Information:
Serial Numbers: 300247414 300247415 300247417 300247418
Distribution Pattern:
CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Voluntary or Mandated:
Voluntary: Firm initiated