Thoratec Corp.: Medical Device Recall in 2020 - (Recall #: Z-0746-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when connected to the HeartMate 3 Left Ventricular Assist System, REF 106524US, UDI or GTIN: 00813024013297, and 106524, UDI or GTIN: 00813024011712 - Product Usage: Provides power to the System Controller and pump. - Provides power to the System Monitor when it is connected to the Power Module. - Connects the System Monitor to the System Controller for monitoring purposes. - Echoes System Controller alarms. The Mobile Power Unit is for home or clinical use when the patient does not require monitoring using the System Monitor. The Mobile Power Unit is used when the patient is indoors, stationary, or sleeping. The System Controller and the Mobile Power Unit are connected through the Mobile Power Unit patient cable. The cable transfers power from the Mobile Power Unit to the System Controller.

Product Classification:

Class II

Date Initiated: December 2, 2019
Date Posted: January 8, 2020
Recall Number: Z-0746-2020
Event ID: 84379
Reason for Recall:

Excessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which could lead to no or reduced blood flow from the left ventricular assist system.

Status: Terminated
Product Quantity: 13871
Code Information:

All serial numbers.

Distribution Pattern:

US Nationwide distribution in the states of MN, PA, IL, NJ, WI, AZ, AR, TN, MO, TX, MA, NE, MI, CA, NC, CO, OH, DE, FL, NY, IN, GA, VA, DC, CT, OK, UT, KY, MD, OR, NM, ME, SC, IA, WA, LA, NV, AL, KS, MS, WV.

Voluntary or Mandated:

Voluntary: Firm initiated