Thoratec Corp.: Medical Device Recall in 2021 - (Recall #: Z-0308-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)

Product Classification:

Class II

Date Initiated: October 22, 2021
Date Posted: December 8, 2021
Recall Number: Z-0308-2022
Event ID: 88876
Reason for Recall:

If LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth connection, the Bluetooth connectivity interference may cause the system App to close or fail to open. Once Bluetooth adaptor is connected, barring disconnection, all functionality is unaffected; the device will continue to operate as expected.

Status: Ongoing
Product Quantity: 988
Code Information:

All serial numbers.

Distribution Pattern:

Worldwide distribution - US Nationwide distribution in the states of FL, IL, PA, AR, GA, WI, AZ, TN, TX, MA, NE, CA, VA, NC, OH, MI, DC, IN, NJ, CT, OK, MD, NY, ME, IA, SC, KY, LA, OR, NM, WA, SD, MO, NV, HI, AL, MS, UT, KS and the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, NETHERLANDS, NEW ZEALAND, POLAND, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, UNITED ARAB EMIRATES, UNITED KINGDOM.

Voluntary or Mandated:

Voluntary: Firm initiated