Topcon Medical Systems, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0952-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained healthcare professionals and clinicians to access archived multimedia data, forms and patient information.
Product Classification:
Class II
Date Initiated: December 18, 2020
Date Posted: February 10, 2021
Recall Number: Z-0952-2021
Event ID: 87086
Reason for Recall:
Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record
Status: Terminated
Product Quantity: 13 units
Code Information:
Software Version 3.1
Distribution Pattern:
AZ, MO, NC, NY, PA, TX
Voluntary or Mandated:
Voluntary: Firm initiated