Torax Medical, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2037-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

LINX Reflux Management System, Implant Suture, 12-17 Bead, 0.7T, Product Code LS-xx, sterile, Rx only. Labeled as: a. Product Code LS-12; b. Product Code LS-13; c. Product Code LS-14; d. Product Code LS-15; e. Product Code LS-16; f. Product Code LS-17; Laparoscopic, fundic-sparing anti-reflux procedure indicated for patients diagnosed with Gastroesophageal Reflux Disease (GERD) as defined by abnormal pH testing, and who are seeking an alternative to continuous acid suppression therapy (i.e. proton pump inhibitors or equivalent) in the management of their GERD.

Product Classification:

Class II

Date Initiated: April 20, 2018
Date Posted: June 6, 2018
Recall Number: Z-2037-2018
Event ID: 79822
Reason for Recall:

Possible out of specification condition which may allow a bead component to separate from an adjacent wire link. This condition may result in a discontinuous or open LINX device.

Status: Terminated
Product Quantity: 9,131 devices in total
Code Information:

a. Product Code LS-12, Part No. 2170-6, GTIN 00855106005011; b. Product Code LS-13, Part No. 2170-7, GTIN 00855106005028; c. Product Code LS-14, Part No. 2170-8, GTIN 00855106005035; d. Product Code LS-15, Part No. 2170-9, GTIN 00855106005042; e. Product Code LS-16, Part No. 2170-10, GTIN 00855106005059; f. Product Code LS-17, Part No. 2170-11, GTIN 00855106005066; Lot No. 6100 through 14055, 15286, 15287, 15314, 15315, 15317, and 16557

Distribution Pattern:

Worldwide Distribution: US (nationwide) in states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY; and in countries of: Austria, France, Germany, Italy, Scotland, Spain, Switzerland, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated