Tornier, Inc: Medical Device Recall in 2013 - (Recall #: Z-2199-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
AEQUALIS Reversed II or AEQUALIS Reversed Fracture Impactor. These products are contained within the Aequalis reversed II Instrument set tray no.YKAD83 and the Aequalis reversed fracture instrument set tray no.YKAD95. The items affected are MWD21, MWD023, MWD024 and MWD025. The impactor is an instrument used during shoulder arthroplasty; it is not implanted. The impactor is used to set the glenoid sphere, poly insert, spacer, and hemi adapter into place during surgery. The impactor is a two piece instrument which consists of a tip and a handle. The tip is made of plastic and metal that is pressed together for assembly.
Class II
Following the identification of the disassociation of an impactor tip during surgery, Tornier is initiating a voluntary recall of all lots of Aequalis Reversed II and Reversed Fracture impactors. Upon review of the event, Tornier determined that a redesign of the instrument was warranted. These products are contained within the Aequalis Revered II instrument set tray no YKAD83 and the Aequalis R
all lots
.Worldwide Distribution - USA including Puerto Rico and the states of AL, AZ, CA, CO,CT, FL, GA, IL, KY, MD, MA, MI, MN, NE, NJ, NC, OH, OR, PA and TX., and the countries of CANADA, MEXICO, HONG KONG, SOUTH AFRICA, ARGENTINA, COLOMBIA, ISRAEL, AUSTRALIA, BELGIUM, FINLAND,FRANCE, GERMANY, ITALY, NETHERLANDS, SLOVENIA, SPAIN, SWITZERLAND and UNITED KINGDOM
Voluntary: Firm initiated