Toshiba American Medical Systems Inc: Medical Device Recall in 2013 - (Recall #: Z-1068-2013)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2013.
Data Source: FDA.
Product Description:
Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.
Product Classification:
Class II
Date Initiated: November 14, 2012
Date Posted: April 17, 2013
Recall Number: Z-1068-2013
Event ID: 64673
Reason for Recall:
Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.
Status: Terminated
Product Quantity: 1 unit
Code Information:
Serial No: 1CB1262006
Distribution Pattern:
Distributed in the state of MD.
Voluntary or Mandated:
Voluntary: Firm initiated